Data Sheet
PDFINFO H 5606- 01
64-Channel Relay Multiplexer
Agilent E1460A
1-Slot, C-size, register based
Armature latching relay channels
Configuration for testing insulation
Includes QUIC easy-to-use terminal block
Numerous multiplexer topologies
Configurable for scanning voltmeter applications
Description
The Agilent E1460A High-Density Relay Multiplexer is a
C-size, 1-slot, register-based VXI module. This
64- channel multiplexer, using latching armature switches,
offers a highly configurable, high point-count switching
topology. Switching topologies include 64 two- wire, 32 three-
wire, 32 four-wire, or 128 single-ended latching relay channels.
This multiplexer consists of a component card with switches
(labeled E1460-66202) and the QUIC screw terminal block
(E1460-80011) that plugs onto the component card.
Use of SCPI commands or status bit jumpers on the terminal
card configures the E1460A “wire mode” as either a 128x1-
wire, 64x2-wire, 32x3-wire, or 32x4- wire multiplexer.
Applications for the E1460A include wire harness and cable
testing, semiconductor testing, and printed circuit board
testing.
Refer to the Agilent Technologies Website for instrument
driver availability and downloading instructions, as well as for
recent product updates, if applicable.
Agilent Technologies
Przeglądanie stron 0
1 2 3 4 5 6 ... 193 194

Podsumowanie treści

Strona 1 - 64-Channel Relay Multiplexer

Data SheetPDFINFO H 5606- 0164-Channel Relay MultiplexerAgilent E1460A• 1-Slot, C-size, register based • Armature latching relay channels • Configurat

Strona 2 - General Specifications

4Chapter 3 - Relay Multiplexer Command Reference ... 59About This Chapter ...

Strona 3 - Cooling/Slot

94 Relay Multiplexer Command Reference Chapter 3 TRIG:SOUR EXT!Select external triggeringSCAN (@100:107)!Scan channels 0 to 7 in bank 0INIT!Begin sc

Strona 4 - PDFINFO H 5 6 0 6 - 0 4

Relay Multiplexer Command Reference 95Chapter 3IEEE 488.2 Common Commands ReferenceThe following table lists the IEEE 488.2

Strona 5 - Ordering Information

96 Relay Multiplexer Command Reference Chapter 3 SCPI Commands Quick ReferenceThe following table summarizes SCPI commands for the E1460A Relay Mult

Strona 6

Relay Multiplexer Specifications 97Appendix AAppendix ARelay Multiplexer Specifications Input CharacteristicsMaximum Voltag

Strona 7 - User’s Manual

98 Relay Multiplexer Specifications Appendix A Notes:

Strona 8

Register-Based Programming 99Appendix BAppendix BRegister-Based Programming About This AppendixThis appendix contains infor

Strona 9

100 Register-Based Programming Appendix B A16 Address Space Outside the Command ModuleWhen the E1406A command module is not part of your VXIbus syst

Strona 10

Register-Based Programming 101Appendix BA16 Address Space Inside the Command Module or MainframeWhen the A16 address space

Strona 11

102 Register-Based Programming Appendix B Register DescriptionsThere are ten WRITE and twelve READ registers on the multiplexer. This section contai

Strona 12

Register-Based Programming 103Appendix BStatus/Control RegisterYou can perform reads and writes to the Status/Control Regis

Strona 13

5 Appendix A - Relay Multiplexer Specifications ... 97Appendix B - Register-Based Programming

Strona 14 - Documentation History

104 Register-Based Programming Appendix B ID and Device Type RegistersID Register: Reading this register returns: FFFF that shows that Hewlett-Pack

Strona 15 - Declaration of Conformity

Register-Based Programming 105Appendix B Bank 2 Relay Control Registerbase +241615 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0Write*

Strona 16

106 Register-Based Programming Appendix B Bank 7 Relay Control Registerbase + 2E1615 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0Write* Undefined CH7 CH6 CH5

Strona 17 - Getting Started

Register-Based Programming 107Appendix BProgramming Examples Some examples follow to illustrate programming the multiplexer

Strona 18 - Switches

108 Register-Based Programming Appendix B Example: Opening/Closing Multiplexer ChannelsThe flowchart in Figure B-3 shows one way to close (or open)

Strona 19

Register-Based Programming 109Appendix B#include <stdio.h>#include <chpib.h>#include <cfunc.h>#define

Strona 20 - Basic Operating

110 Register-Based Programming Appendix B Example: Using a Multimeter with a MultiplexerThe flowchart in Figure B-4 shows a typical timing sequence

Strona 21 - Configuring the Multiplexer

Register-Based Programming 111Appendix BThe multimeter’s Status bit 5 is monitored to determine when the analog- to-digital

Strona 22 - Address Switch

112 Register-Based Programming Appendix B Example: Scanning Channels This C program example is similar to the closing/opening example except that it

Strona 23 - Priority

Register-Based Programming 113Appendix B IOENTER(70900L, &read); bit_number = ((long) (read) >6 & 1); } retu

Strona 25 - Reconfiguring the Relay

114 Register-Based Programming Appendix B /* Open a device session for the E1460A at laddr 112. */ e1460a = iopen(E1460A); /* Map in the

Strona 26

Relay Multiplexer Error Messages 115Appendix CAppendix CRelay Multiplexer Error MessagesTable C-1 lists the error messages

Strona 27 - Multiplexer in a

116 Relay Multiplexer Error Messages Appendix CNotes:

Strona 28 - Analog Bus

Index 117IndexE1460A Relay Multiplexer User’s ManualAA16 address space inside command module, 101outside command module, 100ABORt subsystem, 62ad

Strona 29 - Configuring Terminal Modules

118 Index E (continued)examples (continued)Querying Operating Mode, 76Querying the Scan Port, 83Querying the Scanning Mode, 82Querying Trigger Slo

Strona 30

Index 119Pprogramming the multiplexer, 29RREAD registers, description, 102reconfiguring relay switch, 19register addressing, 99register offset,

Strona 31 - Connecting User

120 Index WWARNINGS, 8warnings, 15warranty statement, 7wire jumper functions, 19wire jumpers, setting, 18wiring terminal modules, 26WRITE regis

Strona 33

122 Index

Strona 34 - Modules to the

Agilent 75000 SERIES CAgilent E1460A64-Channel Relay MultiplexerService ManualEnclosed is the Service Manual for the Agilent E1460A 64-ChannelRelay

Strona 35 - Programming the Multiplexer

7AGILENT TECHNOLOGIES WARRANTY STATEMENTAGILENT PRODUCT: E1460A 64-Channel Relay Multiplexer Module DURATION OF WARRANTY:

Strona 37 - Multiplexer Card

ContentsChapter 1 - General InformationIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 38 - Addresses

Chapter 3 - Replaceable PartsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 39 - Module Channel List

CertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. AgilentTechn

Strona 40

Frame or chassis ground terminal—typicallyconnects to the equipment’s metal frame.Alternating current (AC).Direct current (DC). Indicates hazardous vo

Strona 41 - Using the Relay Multiplexer

DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Agilent E1460A 64-Channel Relay Multiple

Strona 42

Notes6 Agilent E1460A 64-Channel Relay Multiplexer Module Service Manual

Strona 43 - Switching Channels

NotesAgilent E1460A 64-Channel Relay Multiplexer Module Service Manual 7

Strona 44 - Channels (One-Wire)

Notes8 Agilent E1460A 64-Channel Relay Multiplexer Module Service Manual

Strona 45 - Channels (Two-Wire)

Chapter 1General InformationIntroduction This manual contains information required to test, troubleshoot, and repairthe Agilent E1460A relay multiplex

Strona 46 - Channels (Three-Wire)

8Safety SymbolsInstruction manual symbol affixed to product. Indicates that the user must refer to the manual for specific WARNING or CAUTION informat

Strona 47 - Channels (Four-Wire)

Relay Life Electromechanical relays are subject to normal wear-out. Relay lifedepends on several factors. The effects of loading and switching frequ

Strona 48 - SWITCH CARD TERMINAL MODULE

SafetyConsiderationsThis product is a Safety Class I instrument that is provided with a protectiveearth terminal when installed in the mainframe. The

Strona 49 - Scanning Channels

WARNINGUSING AUTOTRANSFORMERS. If the mainframe is to beenergized via an autotransformer (for voltage reduction) makesure the common terminal is conn

Strona 50 - Command Module

IncomingInspectionUse the following steps as guidelines to perform initial (incoming)inspection of the Agilent E1460A.WARNINGTo avoid possible hazardo

Strona 51

ShippingGuidelinesFollow the steps in Figure 1-2 to return the relay multiplexer module to anAgilent Technologies Sales and Support Office or Service

Strona 52 - Multimeter

Environment The recommended operating environment for the multiplexer is:Environment Temperature HumidityOperating 0oC to +55oC <65% relative (0oC

Strona 53 - Channels Using 3457A

Recommended Test EquipmentTable 1-1 lists the test equipment recommended for testing and servicingthe module. Essential requirements for each piece o

Strona 54

Chapter 2Verification TestsIntroduction The two levels of test procedures described in this chapter are used to verifythat the relay multiplexer modul

Strona 55 - Measurements

NoteIf you are using an Agilent E1460A terminal block, and some or all of therelay module jumpers have been removed, you should add the followingconne

Strona 56 - Multimeter Resistance

FunctionalVerificationThe Functional Verification Test for the relay multiplexer module consistsof sending the *TST? command and checking the response

Strona 57 - Complete Bit

9Declaration of ConformityDeclarations of Conformity for this product and for other Agilent products may be downloaded from the Internet. There are tw

Strona 58 - Example: Connecting the

Self-test ErrorCodesTable 2-1 shows the self-test error codes for the multiplexer module. Themeaning of each code is given in the right-hand column.

Strona 59

Agilent E1460APerformanceVerificationThe procedures in this section are used to test the multiplexer’s electricalperformance using the specifications

Strona 60 - TERMINAL MODULE

3. Closed-channel Reading (channel 00)• Send CLOS (@ss00) to the multiplexer to close bank 0, channel 0.• Trigger the DMM and record the reading.4.

Strona 61

Example:Closed-channelResistance TestThis example performs the Closed-channel Resistance Test for the Agilent E1460A. If the switch card number is no

Strona 62 - Using Polling

Test 2-2. Testing Control RelaysThe E1460-66201 relay module has seven control relays (numbered 0990 to 0996). See the Agilent E1460A User’s Manual f

Strona 63 - Using Interrupts

Table 2-2. Guidelines for Testing Control RelaysControl Relay* Test Connections** SCPI Commands***ChannelReferenceDesignator DMM HI DMM LOTo CloseCon

Strona 64 - Synchronizing the

180 DATA "","AB GUARD","BANK 6 COM LO"190 FOR I=1 TO 13200 FOR J=1 TO 3210 READ Chan_data$(I,J)220 NEXT J230 NEXT I240 !

Strona 65 - Command Types

580 GOSUB Close_ch590 PRINT600 PRINT "END OF TESTS"610 CLEAR SCREEN620 I=990630 FOR J=1 TO 12 STEP 2640 PRINT "Channel: ";I;Chan_d

Strona 66 - Variable Commands

Test 2-3: DC Isolation TestThis test verifies that sufficient DC isolation exists between various pointson the relay multiplexer. The DMM used should

Strona 67 - SCPI Commands Reference

High & Lowto Chassis4. Hardware ConnectionsNoteUse any convenient chassis connection. The illustration shows the DMMLO terminal connected to th

Strona 69 - ARM:COUNt

Example: DC IsolationTestThis example performs the DC Isolation Test for the Agilent E1460A. Ifthe switch card number is not 1, change the value in l

Strona 70 - ARM:COUNt?

PerformanceTest RecordTable 2-3 is a form you can copy and use to record performance verificationtest results for the relay multiplexer. Information

Strona 71 - INITiate

Table 2-3. Performance Test Record for the E1460A (Page 1 of 5)Test Facility:Name _________________________________Address____________________________

Strona 72 - INITiate[:IMMediate]

Table 2-3. Performance Test Record for the E1460A (Page 2 of 5)Model __________________________ Report No. _________________ Date ______________

Strona 73 - Enabling a Single Scan

Table 2-3. Performance Test Record for the E1460A (Page 3 of 5)Model _________________________ Report No. __________________________ Date_________

Strona 74 - OUTPut:ECLTrgn[:STATe]

Table 2-3. Performance Test Record for the E1460A (Page 4 of 5)Model _________________________ Report No. ___________________________ Date________

Strona 75 - OUTPut[:EXTernal][:STATe]

Table 2-3. Performance Test Record for the E1460A (Page 5 of 5)Model _________________________ Report No. ___________________________ Date________

Strona 76 - OUTPut:TTLTrgn[:STATe]

Chapter 3Replaceable PartsIntroduction This chapter contains information for ordering replaceable parts for the Agilent E1460A relay multiplexer modul

Strona 77 - OUTPut:TTLTrgn[:STATe]?

Table 3-1. E1400-80011 Terminal Module Case Assembly(for modules with serial numbers greater than 3002A03220)ReferenceDesignatorAgilent PartNumberQty

Strona 78 - [ROUTe:]

Table 3-2. E1400-80001 Terminal Module Case Assembly(for modules with serial numbers less than 3002A03219)ReferenceDesignatorAgilent PartNumberQty Par

Strona 79

Getting Started 11Chapter 1Chapter 1Getting Started Using This ChapterThis chapter describes the E1460A 64-Channel Relay Mu

Strona 80 - [ROUTe:]CLOSe?

Table 3-3. E1460A Terminal Module Replaceable PartsReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberE1460-665101PC ASS

Strona 81 - [ROUTe:]FUNCtion

Table 3-4. Agilent E1460A Option A3E Terminal Module Replaceable PartsReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumbe

Strona 82 - [ROUTe:]FUNCtion?

Table 3-5. ME1460A Relay Module Mechanical Replaceable Parts(for modules with serial numbers greater than 3002A03220)ReferenceDesignatorAgilent PartNu

Strona 83 - Parameters

Table 3-6. E1460-66201 Relay Module Mechanical Replaceable Parts(for modules with serial numbers less than 3002A03219)ReferenceDesignatorAgilent PartN

Strona 84

Table 3-7. ME1460A Component Level Parts ListReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberE1460-665021MULTIPLEXER

Strona 85 - [ROUTe:]SCAN

Table 3-7. ME1460A Component Level Parts List (continued)ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberRP1-RP3 181

Strona 86

Figure 3-7. Relay Module Component Locator46 Replaceable Parts Agilent E1460A Service Manual

Strona 87

Table 3-5. Code List of Manufacturer’sMfr.CodeManufacturer’sNameManufacturer’sAddressZipCode00779 AMP Inc Harrisburg, PA 1711101121 Allen-Bradley Co.

Strona 88 - [ROUTe:]SCAN:PORT

48 Replaceable Parts Agilent E1460A Service Manual

Strona 89 - [ROUTe:]SCAN:PORT?

Chapter 4ServiceIntroduction This chapter contains service information for the Agilent E1460A64-Channel Relay Multiplexer module, including troublesho

Strona 90 - Subsystem Syntax STATus

12 Getting Started Chapter 1 Channel Relay SwitchesThe channel relay switches are separated into eight banks. Each bank has eight switchable channel

Strona 91 - Operation Status Register

Table 4-1. Agilent E1460A Common ProblemsProblemType SymptomPossibleSolutionsSelf-test Errors Non-zero error code in responseto the *TST? command.• Se

Strona 92 - STATus:OPERation:ENABle?

Testing theAssemblyYou can use the tests and checks in Table 4-2 to isolate the problem. SeeFigures 3-1 through 3-4 in Chapter 3 for locations of mec

Strona 93 - STATus:PRESet

Matching Relaysto ChannelsUse Table 4-3 to find the reference designator of any relay on theE1460-66502 relay module. T

Strona 94 - Subsystem Syntax SYSTem

Disassembly Use the following procedures to disassemble the ME1460A relay module(see Figure 4-1):To remove the top shield:• Remove the six T10 Torx sc

Strona 95 - SYSTem:CTYPe?

Repair/MaintenanceGuidelinesThis section provides guidelines for repairing and maintaining the Agilent E1460A multiplexer module, including:• ESD prec

Strona 96 - SYSTem:ERRor?

Component Locators and Schematic DiagramsTable 4-4 lists Component Locator Diagrams and Schematic Diagrams forthe Agilent E1460A Relay Multiplexer mod

Strona 97 - TRIGger[:IMMediate]

56 Service Agilent E1460A Service Manual

Strona 98 - TRIGger:SOURce

Appendix AVerification Tests - C ProgramsNoteThese examples assume a Multiplexer logical address setting of 70914. Ifyour Multiplexer has a different

Strona 99

PerformanceVerificationTestsThese programs are designed to do the Performance Verification Testsfound in Chapter 2 - Verification Tests.Test 2-1: Clos

Strona 100 - TRIGger:SOURce?

printf ("\n\nHigh to Common Measurements");printf ("\n\n 1. Connect DMM Input HI lead to High lines");printf ("\n 2. Connect

Strona 101 - <mask>

Getting Started 13Chapter 1Banks are arranged as follows:• Bank 0 includes channels 00 through 07 and COM• Bank 1 includes

Strona 102 - SCPI Commands Quick Reference

Test 2-2: TestingControl RelaysThis program performs the control relays test found in Chapter 2 -Verification Tests. See Chapter 2 for a detailed des

Strona 103 - Appendix A

printf ("\n\nTesting Channel %u %s", channel, data[i][0]); printf ("\n Connect DVM HI to %s", data[i][1]); printf ("

Strona 104

Test 2-3: DCIsolationThis test verifies that sufficient DC isolation exists between various pointson the multiplexer. See Chapter 2 for hardware conn

Strona 105 - Register-Based Programming

printf ("\n\nConnect DMM HI and LO to E1460A High and Low lines");getchar ();ipromptf (dm, "TRIG SGL\n", "%t", reading);

Strona 106 - Outside the Command

64 Example C Programs Agilent E1460A Service Manual

Strona 107 - Inside the Command

PDFINFO H 5 6 0 6 - 0 22ConfigurationThe switch consists of eight banks of eight Hi and Loswitches, each bank having its own eight Hi and Lo common.Th

Strona 108 - Register Descriptions

Table 1-1. Control Relay FunctionsConnects Cable Test or one-wire LO REF terminal to the one-wire LO COM terminal.Connects upper 32 channels (banks 4

Strona 109 - Register

Getting Started 15Chapter 1Configuring the MultiplexerThis section gives guidelines to configure the relay switch card. See

Strona 110 - Relay Control

16 Getting Started Chapter 1 Setting the Logical Address SwitchPlug-in modules installed in an mainframe or used with a command module are treated a

Strona 111

Getting Started 17Chapter 1Setting the Status Register SwitchFour bits of the Status Register (bits 10-13) define whether t

Strona 112

18 Getting Started Chapter 1 NOTE The interrupt priority jumper MUST be installed in position 1 when using the E1406 command module. Level X interru

Strona 113 - Programming Examples

Getting Started 19Chapter 1Wire Jumper Functions With the exception of JM1, wire jumpers are changed in pairs. Functions of

Strona 114 - Multiplexer Channels

20 Getting Started Chapter 1 NOTE When wire jumpers JM10 through JM17 are removed, the odd-numbered banks can no longer be connected to the analog b

Strona 115

Getting Started 21Chapter 1Installing the Multiplexer in a MainframeThe E1460A can be installed in any slot (except slot 0)

Strona 116 - Multiplexer

22 Getting Started Chapter 1 Connecting the Analog BusFigure 1-7 shows how to connect the analog bus between multiple multiplexer modules and to the

Strona 117 - and Status Registers

Getting Started 23Chapter 1Configuring Terminal ModulesThe E1460A 64-Channel Relay Multiplexer consists of a relay switch c

Strona 118 - Channels

PDFINFO H 5 6 0 6 - 0 33Instrument DriversSee the Agilent Technologies Website (http://www.agilent.com/find/inst_drivers) for driver availability and

Strona 119 - Channels (HP-UX)

Table 1-3. Option A3E Terminal Module AccessoriesCrimp-and-Insert ContactsThe crimp-and-insert connector is normally supplied with Option A3E. Contac

Strona 120

Getting Started 25Chapter 1Connecting User InputsFigure 1-10 shows the front panel of the E1460A and the multiplexer’s conn

Strona 121 - Appendix C

26 Getting Started Chapter 1 Wiring Terminal ModulesFigures 1-11 and 1-12 show suggested steps to connect field wiring (user inputs) to a terminal m

Strona 122

Getting Started 27Chapter 1Figure 1-12. Steps to Wire Terminal Modules (continued) Terminal Module onto the Multiplexer

Strona 123

28 Getting Started Chapter 1 Attaching Terminal Modules to the MultiplexerFigure 1-13 shows how to attach a terminal module to the multiplexer and h

Strona 124

Getting Started 29Chapter 1Programming the MultiplexerThe multiplexer modules are programmed using either a switchbox or sc

Strona 125

30 Getting Started Chapter 1 SYSTEM,E1406A,A.08.00,ROM;IBASIC,IBASIC,A.O4.02,ROM;VOLTMTR,E1326B,A.06.00,ROM;SWITCH,SWITCHBOX,A.07.00,ROM; COUNTER,

Strona 126

Getting Started 31Chapter 1Multiplexer Card NumbersThe multiplexer card number identifies the module within a switchbox or

Strona 127 - Index 121

32 Getting Started Chapter 1 Multiplexer Channel AddressesFor the E1460A, the channel address (channel_list) has the form:(@ssbc) for two-wire, thre

Strona 128 - 122 Index

Getting Started 33Chapter 1or any combination of the above.Low or High Terminal NumberThe LO or HI terminal number is speci

Strona 129 - *E1460-90012*

PDFINFO H 5 6 0 6 - 0 44Agilent E1460A with MUX-to-MUX and MUX-to-multimeter analog bus cabling

Strona 130

34 Getting Started Chapter 1 multiplexerInitial Operation You can use the following example program to verify initial multiplexer module operation b

Strona 131 - Contents

Using the Relay Multiplexer 35Chapter 2Chapter 2Using the Relay MultiplexerUsing This Chapter This chapter shows how to use

Strona 132

36 Using the Relay Multiplexer Chapter 2Table 2-2 summarizes the query commands you can use to determine the configuration or state of the multiplex

Strona 133

Using the Relay Multiplexer 37Chapter 2Switching ChannelsFor general purpose switching, you can switch channels (connect or

Strona 134 - Safety Symbols

38 Using the Relay Multiplexer Chapter 2Analog Bus Connection When Not Scanning. When opening and closing individual channels in all four modes of o

Strona 135 - DECLARATION OF CONFORMITY

Using the Relay Multiplexer 39Chapter 2Example: Switching Channels (Two-Wire)This example illustrates two-wire mode operati

Strona 136

40 Using the Relay Multiplexer Chapter 2To connect the HI and LO terminals of bank 0 channels 0 and 7 to the bank 0 COM terminals, execute: FUNC 1,W

Strona 137

Using the Relay Multiplexer 41Chapter 2To connect the HI and LO terminals of bank 0 channel 0 and the LO terminal of bank 4

Strona 138

42 Using the Relay Multiplexer Chapter 2To connect the HI and LO terminals of bank 0 channel 0 and the HI and LO terminals of bank 4 channel 0 to th

Strona 139 - General Information

Using the Relay Multiplexer 43Chapter 2Scanning ChannelsScanning the multiplexer module channels consists of closing bank c

Strona 140

PDFINFO H 5606- 055Ordering InformationExtra Screw Terminal Block E1460-80011Extra Crimp-and-Insert Terminal Block(if ordered separately)* E1460-80012

Strona 141 - Considerations

44 Using the Relay Multiplexer Chapter 2Scanning Requirements of a Switchbox (With a Command Module). To scan modules in a switchbox, you must nnow

Strona 142

Using the Relay Multiplexer 45Chapter 210 DIM Readings(1:64)!Dimensions computer to store readings20 OUTPUT 70903;"

Strona 143 - Inspection

46 Using the Relay Multiplexer Chapter 2120 OUTPUT 70914;"TRIG:SOUR TTLT1"!Multiplexer to advance scan when trigger received in TTL trigge

Strona 144 - Guidelines

Using the Relay Multiplexer 47Chapter 210 ASSIGN @Dvm TO 7090320 ASSIGN @Mux TO 7091430 DIM A$[80],Rdgs(1:64)40

Strona 145 - Description

48 Using the Relay Multiplexer Chapter 210 OUTPUT 722;"TRIG EXT;DCV;MEM FIFO"!Configures the 3457A multimeter to external trigger to mea

Strona 146 - Test Equipment

Using the Relay Multiplexer 49Chapter 2Example: Scanning Multimeter DCV MeasurementsThis example uses the E1406A Command Mo

Strona 147 - Verification Tests

50 Using the Relay Multiplexer Chapter 2Example: Scanning Multimeter Resistance Measurements This program uses an E1406A Command Module to verify th

Strona 148

Using the Relay Multiplexer 51Chapter 2Miscellaneous Multiplexer FunctionsThis section describes some miscellaneous multipl

Strona 149 - Verification

52 Using the Relay Multiplexer Chapter 2Using the Analog BusThe multiplexer can be configured to perform voltage, two-wire ohm, or four-wire ohm mea

Strona 150 - Self-test Error

Using the Relay Multiplexer 53Chapter 2Example: Cable Testing You can connect and test multi-conductor cables or wiring har

Strona 151 - Performance

Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive, while minimiz

Strona 152

54 Using the Relay Multiplexer Chapter 2For this example, the multiplexer module is configured to test a 4-conductor cable. Jumper JM1 must be remov

Strona 153 - Resistance Test

Using the Relay Multiplexer 55Chapter 210 OUTPUT 70903;"*RST"20 OUTPUT 70903;"CONF:RES;*OPC?"30 ENTER

Strona 154

56 Using the Relay Multiplexer Chapter 2Saving and Recalling StatesThe *SAV <numeric_state> command saves the current instrument state. The st

Strona 155 - Control Relay Test

Using the Relay Multiplexer 57Chapter 2Example: Error Checking Using InterruptsThe second approach to error checking involv

Strona 156

58 Using the Relay Multiplexer Chapter 2Synchronizing the MultiplexerThis section discusses synchronizing the multiplexer module to other instrument

Strona 157

Relay Multiplexer Command Reference 59Chapter 3Chapter 3Relay Multiplexer Command ReferenceAbout This ChapterThis chapter d

Strona 158 - Test 2-3: DC Isolation Test

60 Relay Multiplexer Command Reference Chapter 3 Command Separator A colon (:) always separates one command from the next lower-level command, such

Strona 159 - High & Low

Relay Multiplexer Command Reference 61Chapter 3Parameter Types The following table contains explanations and examples of pa

Strona 160 - Example: DC Isolation

62 Relay Multiplexer Command Reference Chapter 3 ABORtThe ABORt command stops a scan in progress when the scan is enabled via the interface and the

Strona 161 - Test Record

Relay Multiplexer Command Reference 63Chapter 3ARMThe ARM subsystem selects the number of scanning cycles (1 to 32767) for

Strona 162

Agilent TechnologiesE1460A 64-ChannelRelay Multiplexer ModuleUser’s Manual

Strona 163 - Test Equipment Used:

64 Relay Multiplexer Command Reference Chapter 3 ARM:COUNt?ARM:COUNt? [MIN I MAX] returns the current number of scanning cycles set by ARM:COUNt. Th

Strona 164

Relay Multiplexer Command Reference 65Chapter 3INITiateThe INITiate command subsystem selects continuous scanning cycles an

Strona 165

66 Relay Multiplexer Command Reference Chapter 3 Example Enabling Continuous ScanningThis example enables continuous scanning of bank 3, channels 0

Strona 166 - High & Low to Chassis

Relay Multiplexer Command Reference 67Chapter 3Example Enabling a Single ScanThis example enables a single scan of channels

Strona 167 - Replaceable Parts

68 Relay Multiplexer Command Reference Chapter 3 OUTPutThe OUTPut command subsystem selects the source of the output trigger generated when a channe

Strona 168

Relay Multiplexer Command Reference 69Chapter 3One Output Selected at a Time: Only one output (ECLTrg0 or 1; TTLTrg0, 1, 2,

Strona 169

70 Relay Multiplexer Command Reference Chapter 3 Comments Enabling “Trig Out” Port: When enabled, a pulse is output from the “Trig Out” port after e

Strona 170

Relay Multiplexer Command Reference 71Chapter 3Parameters Comments Enabling TTL Trigger Bus: When enabled, a negative-going

Strona 171

72 Relay Multiplexer Command Reference Chapter 3 [ROUTe:]The [ROUTe:] command subsystem controls switching and scanning operations for multiplexer m

Strona 172

Relay Multiplexer Command Reference 73Chapter 3• WIRE2: Configures the E1460A as two independent 2x32 multiplexers.• WIRE2X

Strona 174 - (Continued on next page)

74 Relay Multiplexer Command Reference Chapter 3 Related Commands: [ROUTe:]OPEN, CLOSe?, SCAN*RST Condition: All multiplexer channels are open.Examp

Strona 175

Relay Multiplexer Command Reference 75Chapter 3[ROUTe:]FUNCtion[ROUTe:]FUNCtion <card_number>, <function> selec

Strona 176

76 Relay Multiplexer Command Reference Chapter 3 When closing a channel in three-wire mode, only the lower bank (0-3) is specified and the upper ban

Strona 177

Relay Multiplexer Command Reference 77Chapter 3[ROUTe:]OPEN[ROUTe:]OPEN <channel_list> opens the multiplexer channels

Strona 178

78 Relay Multiplexer Command Reference Chapter 3 NOTE Channel numbers can be in the channel_list in any random order. Hovever, opening order for mul

Strona 179 - Chapter 4

Relay Multiplexer Command Reference 79Chapter 3[ROUTe:]OPEN?[ROUTe:]OPEN? <channel_list> returns the current state of

Strona 180

80 Relay Multiplexer Command Reference Chapter 3 Comment Defining Scan List: When ROUTe:SCAN is executed, the channel list is checked for valid card

Strona 181 - Assembly

Relay Multiplexer Command Reference 81Chapter 3Comments Order of Command Execution: The [ROUTe:]SCAN:MODE and [ROUTe:]FUNCt

Strona 182 - Matching Relays

82 Relay Multiplexer Command Reference Chapter 3 [ROUTe:]SCAN:MODE?[ROUTe:]SCAN:MODE? returns the current state of the scan mode. Comments Values Re

Strona 183

Relay Multiplexer Command Reference 83Chapter 3Example Selecting the Analog Bus PortThis example selects the four-wire ohms

Strona 184 - Maintenance

3 ContentsE1460A Relay Multiplexer User’s ManualFront Matter...

Strona 185 - Diagrams

84 Relay Multiplexer Command Reference Chapter 3 STATusThe STATus subsystem reports the bit values of the Operation Status Register. It also allows

Strona 186

Relay Multiplexer Command Reference 85Chapter 3Figure 3-1. Relay Multiplexer Status System Registers0123456789101112131415

Strona 187

86 Relay Multiplexer Command Reference Chapter 3 STATus:OPERation:CONDition?STATus:OPERation:CONDition? returns the state of the Condition Register

Strona 188

Relay Multiplexer Command Reference 87Chapter 3STATus:OPERation[:EVENt]?STATus:OPERation[:EVENt]? returns which bits in the

Strona 189

88 Relay Multiplexer Command Reference Chapter 3 SYSTemThe SYSTem subsystem returns the numbers and messages in the error queue of a switchbox, and

Strona 190 - Control Relays

Relay Multiplexer Command Reference 89Chapter 3SYSTem:CPONSYSTem:CPON <number> | ALL sets the selected module (card)

Strona 191

90 Relay Multiplexer Command Reference Chapter 3 SYSTem:ERRor?SYSTem:ERRor? returns the error numbers and corresponding error messages in the error

Strona 192 - Isolation

Relay Multiplexer Command Reference 91Chapter 3TRIGgerThe TRIGger command subsystem controls the triggering operation of mu

Strona 193

92 Relay Multiplexer Command Reference Chapter 3 TRIGger:SLOPeTRIGger:SLOPe <slope> is used to select the polarity of the output trigger. For

Strona 194

Relay Multiplexer Command Reference 93Chapter 3Using the TRIGger Command: You can use TRIGger[:IMMediate] to advance the sc

Komentarze do niniejszej Instrukcji

Brak uwag